FDA recall record

ICU Medical Product Description/Item: ChemoLock Vial Spike, 20mm/011-CL-80S recall — FDA Z-1682-2022

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the…

Class IIclassification
Z-1682-2022official record ID
Sep 14, 2022report date

Recall record at a glance

Official recordZ-1682-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmICU Medical, Inc.
Report dateSep 14, 2022
Recall initiation dateJun 27, 2022
Quantity358,975
LocationSan Clemente, CA, United States

Official product description

Product Description/Item: ChemoLock Vial Spike, 20mm/011-CL-80S; ChemoLock Bag Spike/CL-10; ChemoLock w/Mini Bag Spike/CL-17; 8 IN (20cm) APPX 1.0ml, EXT SET, MicroClave Clear, ChemoLock PORT, YCONN/CL3950; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger/CL3951; 7.5" (19 cm) Appx 1.5 ml, Ext Set w/ChemoLock Port, Clamp, Graduated Adapter/CL3967; 6 IN(15cm)APPX 0.82ml,EXT SET,MicroClave Clear,ChemoLock PORT,YCONN, ROTATING LUER/CL4118; ChemoLock Kit w/ 5" (13 cm) Bag Spike Adapter w/ChemoLock w/Red Cap, Vented Cap, ChemoLock Port Bag Spike/CL4158; ChemosafeLock Bag Spike/KL-BS001U3; ChemosafeLock Vial Adapter/KL-VA201U3; ChemosafeLock Vial Adapter/KL-VA202U3; ChemosafeLock Vial Adapter/KL-VA321U3; ChemosafeLock Vial Adapter/KL-VA322U3; ChemoLock Vented Vial Spike/011-CL-51; Transfer Set, PUR Yellow w/ChemoLock Port, Spiros/011-CL4155; Puerto ChemoLock/034-CL2100; ChemoLock, Punzon Univ. Vial c/ Filtro, fijacion/034-CL-70; ChemoLock, Punzon Vial Tritan c/ Filtro, 13mm/034-CL-72

Why the product was recalled

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

Distribution information

Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceSep 23, 2026

ICU Medical Plum administration sets recall — FDA Z-3156-2026

Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the…

Class IIFDA: OngoingICU Medical, Inc.