FDA recall record

ICU Medical Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10 recall — FDA Z-1681-2022

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the…

Class IIclassification
Z-1681-2022official record ID
Sep 14, 2022report date

Recall record at a glance

Official recordZ-1681-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmICU Medical, Inc.
Report dateSep 14, 2022
Recall initiation dateJun 27, 2022
Quantity454,635
LocationSan Clemente, CA, United States

Official product description

Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX 1.5ml,ChemoLock BAG SPIKE WITH ADDITIVE PORT/CL-13; ChemoLock Port/CL2100; ChemoLock Port, 5 Units/CL2100-5; ChemoLock Port/CL2100T; Clave Bag Spike w ChemoLock Port/CL2150; 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger/CL3011; 9 IN(23cm)APPX 2.7ml,EXT SET,ChemoLock PORT,2 CLAMPS,2 GRADUATED CONN/CL3364; Syringe Transfer Set w Clave, ChemoLock Port/CL-34; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511T; 30" (76cm) Appx 6.3 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, 0.2 Micron Filter, ChemoLock w Red Cap, Hanger/CL3528; 7" (18 cm) Appx .93 ml, Ext Set w/ChemoLock, Y-Connector, Rotating Luer/CL3535T; Oncology Kit w/5" (13 cm) Add-On Set w ChemoLock Additive Port, Vented Cap, ChemoLock, Spinning Spiros w Red Cap/CL3538; Graduated Connector w ChemoLock Port/CL3900; ONCOLOGY KIT, EXT,ChemoLock PORT, GRAD CONN,TRANSFER,ChemoLock,VIAL SPIKE,ChemoLock PORT/CL3927; Oncology Kit w 5" (13 cm) Add-On Set w ChemoLock w Red Cap, Vented Cap, ChemoLock Port w Bag Spike, Spiros w Red Cap/CL3946; Oncology Kit w 17" (43 cm) Bifuse Ext Set w/ChemoLock w Red Cap, ChemoLock Port, ChemoLock Vented Vial Spike, 13mm ChemoLock/CL3947; Oncology Kit w ChemoLock Port Closed Vial Spike, ChemoLock, ChemoLock Port Bag Spike/CL3952; APPX 0.5ml, ChemoLock BAG SPIKE WITH ADDITIVE PORT,ChemoLock Port/CL3955;

Why the product was recalled

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

Distribution information

Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.