Recall record at a glance
| Official record | Z-2079-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | TEI Biosciences, Inc. |
| Report date | Jul 12, 2023 |
| Recall initiation date | May 23, 2023 |
| Quantity | 33,277 units |
| Location | Boston, MA, United States |
Official product description
PriMatrix Meshed; Dermal Repair Scaffold; Rx Only; Sizes 20x25 cm (1:1), 3x3 cm (2:1), 4x4 cm (2:1), 5x5 cm (2:1), 6x6 cm (2:1), 8x12 cm (2:1), and 8x8 cm (2:1).
Why the product was recalled
Possible out of specification endotoxin test results due to issues with in-process and finished goods endotoxin testing.
Distribution information
Domestic: Nationwide Distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.