FDA recall record

Volcano Prestige Plus 185cm J-Tip: Part number: 9185J recall — FDA Z-1845-2015

During a relabeling procedure at the manufacturing facility, the products were labeled with a shelf life in accordance with the date of the relabeling and…

Class IIclassification
Z-1845-2015official record ID
Jul 1, 2015report date

Recall record at a glance

Official recordZ-1845-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmVolcano Corporation
Report dateJul 1, 2015
Recall initiation dateJun 4, 2015
Quantity162 total (both devices)
LocationRancho Cordova, CA, United States

Official product description

Prestige Plus 185cm J-Tip: Part number: 9185J; Product Usage: The Prestige Plus Wire is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.

Why the product was recalled

During a relabeling procedure at the manufacturing facility, the products were labeled with a shelf life in accordance with the date of the relabeling and not in accordance with the original shelf life date at the time of manufacture

Distribution information

US Nationwide Distribution and one account in Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records