FDA recall record

Merit Medical Systems Prelude Roadster Guide Sheath REF: -PG4F45S -PG4F90S -PG5F45MP recall — FDA Z-2218-2024

Due to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause…

Class IIclassification
Z-2218-2024official record ID
Jul 3, 2024report date

Recall record at a glance

Official recordZ-2218-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMerit Medical Systems, Inc.
Report dateJul 3, 2024
Recall initiation dateMay 22, 2024
Quantity52,410 (updated 06/14/2024)
LocationSouth Jordan, UT, United States

Official product description

Prelude Roadster Guide Sheath REF: -PG4F45S -PG4F90S -PG5F45MP -PG5F45R -PG5F45S -PG5F45S018 -PG5F90MP -PG5F90S -PG5F90S018 -PG6F45HS -PG6F45MP -PG6F45R -PG6F45S -PG6F45S018 -PG6F65MP -PG6F65S -PG6F65SX -PG6F90MP -PG6F90MPX -PG6F90SX -PG6F90SX018 -PG7F45MP -PG7F45MPX -PG7F45RX -PG7F45S -PG7F45SX -PG7F65SX -PG7F90SX -PG8F45MPX -PG8F45S -PG8F45SX -PG8F65SX -PG8F90SX -X9PG700001 The Prelude Guide Sheath Introducer is indicated to be used for the introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.

Why the product was recalled

Due to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause users to over-tightening the luer which may result in the swivel nut detaching from the HVA body.

Distribution information

Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT. NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Brazil, Canada, Cayman Island, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Hong Kong, Republic of Korea, Mexico, New Zealand, Peru, South Africa, and Uruguay.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.