Official product description
Precedence 16 Model # 882350 Product Usage: Precedence (cleared as Griffin) is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. Precedence may produce non-attenuation corrected and attenuation corrected images of the distribution of radiopharmaceuticals in the body as well as x-ray transmission images. The CT transmission data may be used to produce attenuation corrected nuclear medicine images. The nuclear medicine images and the CT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide combined single photon and anatomical data for anatomical localization of the nuclear medicine data. Precedence may be used either as a separate single photon system, a separate CT system or as a combined CT and single photon system. The nuclear medicine and CT images may be transferred to other systems such as a radiation therapy planning system. The Precedence Imaging System should only be used by trained healthcare professionals.
Why the product was recalled
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.
Distribution information
Worldwide Distribution - US Nationwide
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.