FDA recall record

Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967 recall — FDA Z-1321-2024

Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length…

Class IIclassification
Z-1321-2024official record ID
Mar 27, 2024report date

Recall record at a glance

Official recordZ-1321-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPreat Corp
Report dateMar 27, 2024
Recall initiation dateFeb 9, 2024
Quantity561 units
LocationSanta Maria, CA, United States

Official product description

Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967

Why the product was recalled

Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.

Distribution information

U.S. Nationwide distribution in the states of AZ, OH and PA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 16, 2025

Preat recall — FDA Z-2107-2025

Due to manufacturing error, digital analog rotational feature is 180 degrees off.

Class IIFDA: OngoingPreat Corp