FDA recall record

Terumo Cardiovascular Systems Preamendments device-Sarns Dual Cooler-Heater recall — FDA Z-2210-2016

Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the…

Class IIclassification
Z-2210-2016official record ID
Jul 27, 2016report date

Recall record at a glance

Official recordZ-2210-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTerumo Cardiovascular Systems Corporation
Report dateJul 27, 2016
Recall initiation dateJun 8, 2016
Quantity6412 units
LocationAnn Arbor, MI, United States

Official product description

Preamendments device-Sarns Dual Cooler-Heater. The Sarns Dual Cooler-Heater is a self-contained unit for supplying cold and warm water to heat exchangers during open heart surgery. It supplies water at a rate of up to 6.5 gallons per minute (25 liters per minute) at zero head, and will provide a maximum pressure output of 11 pounds per square inch (76,000 pascals). The Sarns TM Dual Cooler-Heater can be used with heat exchangers capable of accepting these flows and pressures. Product Usage: The Sarns Dual Cooler-Heater is indicated for use as a water supply system to a blood heat exchanger in open-heart surgery.

Why the product was recalled

Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Inter chlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.

Distribution information

Domestic: AK, AL, AR, AZ, CA , CO, CT, DC , DE , FL, GA, HI, IA, ID, IL, IN, KS, KY,LA , MA, MD, ME, MI , MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV , NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY Foreign: Argentina, Australia, Belgium,Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, England, Finland, France, Germany Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Lebanon, Malaysia Malta, Mexico, Netherlands, New Zealand, Norway, Philippines, Portugal, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates , Uruguay , Venezuela, Vietnam

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.