Recall record at a glance
| Official record | Z-2663-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Becton Dickinson & Company |
| Report date | Jul 29, 2020 |
| Recall initiation date | Jun 19, 2020 |
| Quantity | 5,296,132 total |
| Location | Franklin Lakes, NJ, United States |
Official product description
PowerLoc MAX Power Injectable Infusion Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers 2132010G (UDI 00801741095504), 2132015G (UDI 00801741095511), 2141910G (UDI 00801741095542), 2142010G (UDI 00801741085680), 2142015G (UDI 00801741085680), 2142210G (UDI 00801741085680)
Why the product was recalled
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
Distribution information
Worldwide Distribution
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.