FDA recall record

Heartware Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU recall — FDA Z-1337-2021

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to…

Class Iclassification
Z-1337-2021official record ID
Apr 14, 2021report date

Recall record at a glance

Official recordZ-1337-2021
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmHeartware, Inc.
Report dateApr 14, 2021
Recall initiation dateFeb 26, 2021
Quantity17992 units (8054 US)
LocationMiami Lakes, FL, United States

Official product description

Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US 2. Medtronic HVAD Controller DC Adapter, Product Number: 1440

Why the product was recalled

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.

Distribution information

Worldwide

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: N/A.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 13, 2023

Heartware PUMP 1103 HVAD recall — FDA Z-0478-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware HVAD recall — FDA Z-0482-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP 1104 HVAD recall — FDA Z-0481-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware HVAD recall — FDA Z-0475-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP 1104 HVAD recall — FDA Z-0479-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP MCS1705PU HVAD recall — FDA Z-0483-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.