FDA recall record

Terex Utilities Power Over Fiber Base Module recall — FDA Z-2745-2026

Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although…

Class IIclassification
Z-2745-2026official record ID
Aug 5, 2026report date

Recall record at a glance

Official recordZ-2745-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmTerex Utilities
Report dateAug 5, 2026
Recall initiation dateMay 11, 2026
Quantity101
LocationWatertown, SD, United States

Official product description

Power Over Fiber Base Module

Why the product was recalled

Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning statement

Distribution information

U.S. Nationwide Distribution and International Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: FDA Mandated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.