FDA recall record

Breg Inc., An Orthofix Company Polar Dressing, Knee/MU/Shldr Intelli-Flo-Breg, PN 10630 Used recall — FDA Z-2326-2012

This recall has been initiated due to confirmation that the Polar Dressing, Knee/MU/Shldr Intelli-Flo-Breg does not meet the required thickness as…

Class IIclassification
Z-2326-2012official record ID
Sep 12, 2012report date

Recall record at a glance

Official recordZ-2326-2012
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBreg Inc., An Orthofix Company
Report dateSep 12, 2012
Recall initiation dateAug 11, 2010
Quantity502 units
LocationCarlsbad, CA, United States

Official product description

Polar Dressing, Knee/MU/Shldr Intelli-Flo-Breg, PN 10630 Used with cold therapy pads to provide insulation barrier between the pad and patient's skin.

Why the product was recalled

This recall has been initiated due to confirmation that the Polar Dressing, Knee/MU/Shldr Intelli-Flo-Breg does not meet the required thickness as specified. The dressing in the sterile pouch measures a thickness of 0.010" instead of 0.020". This change in thickness makes the dressing thinner, less durable and subject to tearing. The product is mislabeled.

Distribution information

Worldwide distribution: USA (nationwide) and country of: Nova Scotia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.