FDA recall record

DePuy Orthopaedics Pinnacle Cups, Part numbers 121701048 121701050 121701052 121701054 recall — FDA Z-0802-2021

Certain Pinnacle Cup devices may potentially exhibit an oversized "minor diameter", which could lead the Apex HE to thread through the shell of the cup…

Class IIclassification
Z-0802-2021official record ID
Jan 13, 2021report date

Recall record at a glance

Official recordZ-0802-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmDePuy Orthopaedics, Inc.
Report dateJan 13, 2021
Recall initiation dateDec 15, 2020
Quantity14161 lots (10 units/lot)
LocationWarsaw, IN, United States

Official product description

Pinnacle Cups, Part numbers 121701048 121701050 121701052 121701054 121701056 121701058 121701060 121701062 121711048 121711052 121712050 121712052 121712056 121722048 121722050 121722052 121722054 121722056 121722058 121722060 121722062 121731048 121731050 121731052 121731054 121731056 121731058 121731060 121731062 121732048 121732050 121732052 121732054 121732056 121732058 121732060 121732062 121732064 - Product Usage: The PINNACLE CUP devices are part of the Pinnacle Hip Solution for the anatomic reconstruction of the hip joint, which promotes prosthetic joint load and function.

Why the product was recalled

Certain Pinnacle Cup devices may potentially exhibit an oversized "minor diameter", which could lead the Apex HE to thread through the shell of the cup without stopping or to protrude internally as a result of "cross-threading". If cross-threading occurs, physicians will not feel the positive stop of the Apex HE against the cup.

Distribution information

Worldwide distribution - US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.