FDA recall record

Richard Wolf GmbH PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial recall — FDA Z-1753-2025

The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce…

Class IIclassification
Z-1753-2025official record ID
May 21, 2025report date

Recall record at a glance

Official recordZ-1753-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmRichard Wolf GmbH
Report dateMay 21, 2025
Recall initiation dateApr 9, 2025
Quantity31 units
LocationKnittlingen, Germany

Official product description

PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with piezoelectric shockwaves (Therapeutic massager).

Why the product was recalled

The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.

Distribution information

US State: GA

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.