FDA recall record

Reichert PHOROPTOR VRX DIGITAL REFRACTION SYSTEM-Digital refractor intended for the recall — FDA Z-2512-2023

Phoroptor VRx Head may detach from the bracket that connects the Phoroptor VRx Head to an ophthalmic stand. If the Phoroptor VRx Head detaches from the…

Class IIclassification
Z-2512-2023official record ID
Sep 6, 2023report date

Recall record at a glance

Official recordZ-2512-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmReichert, Inc.
Report dateSep 6, 2023
Recall initiation dateJul 20, 2023
Quantity53 units; Extended 34 units
LocationDepew, NY, United States

Official product description

PHOROPTOR VRX DIGITAL REFRACTION SYSTEM-Digital refractor intended for the subjective measurement of the refractive error of the eye. MODEL NUMBER: 16242 [Model 16241 is the base model. Model 16242 includes an additional prism feature used for the measurement of ocular deviation in patients with latent or manifest strabismus].

Why the product was recalled

Phoroptor VRx Head may detach from the bracket that connects the Phoroptor VRx Head to an ophthalmic stand. If the Phoroptor VRx Head detaches from the bracket, it can fall onto a patient or device operator

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records