FDA recall record

Philips Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in recall — FDA Z-1801-2020

No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws…

Class IIclassification
Z-1801-2020official record ID
May 6, 2020report date

Recall record at a glance

Official recordZ-1801-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPhilips North America, LLC
Report dateMay 6, 2020
Recall initiation dateMar 5, 2020
Quantity162 units
LocationAndover, MA, United States

Official product description

Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.

Why the product was recalled

No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose..

Distribution information

Worldwide distribution: US (nationwide) including the following states: Arkansas, Arizona, California, Colorado, Delaware, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, Texas, Virginia, Wisconsin. And the following Countries of: Australia, Azerbaijan, Chile, Croatia, Czechoslovakia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Korea South, Lebanon, Mexico, Netherlands, Norway, Pakistan, Poland, Russian Federate, Senegal, Spain, South Africa, Sweden, Switzerland, Taiwan, and Uruguay.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.