Respironics California Philips V60 Ventilators with Power Management PCBA part number recall — FDA Z-1622-2020

Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger…

Class IIclassification
Z-1622-2020official record ID
Apr 8, 2020report date

Recall record at a glance

Official recordZ-1622-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmRespironics California, LLC
Report dateApr 8, 2020
Recall initiation dateMar 5, 2020
Quantity23581
LocationCarlsbad, CA, United States

Official product description

Philips V60 Ventilators with Power Management PCBA part number 1055906

Why the product was recalled

Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia

Distribution information

U.S.: CO, PA, AZ, OH, MT, AR, FL, NC, KS, IL, CA, HI, MD, SC, WY, OR, NY, NE, OK, MN, UT, ND, LA, IN, AL, AK, WV, GA, WI, MI, VA, NJ, TX, WA, SD, KY, TN, MS, MO, DE, MA, VT, CT, IA, NV, ID, ME, NM, DC, NH, VI, RI, PR. O.U.S.: Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Gibraltar, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Jersey, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Malta, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.