Philips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B recall — FDA Z-0882-2023

The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is…

Class Iclassification
Z-0882-2023official record ID
Jan 18, 2023report date

Recall record at a glance

Official recordZ-0882-2023
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmPhilips Respironics, Inc.
Report dateJan 18, 2023
Recall initiation dateDec 7, 2022
Quantity20,807
LocationMurrysville, PA, United States

Official product description

Philips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B, 1054260B, CA1054096B, BT1054260, 1054096B, R1054260, RSL1054260, CA1054096, R1054260B, R1054655TPV, AU1054096, AU1054096B, 1054096B, 1054097, BR1054096, LA1054096, CN1054096, KO1054096, KR1054096, R1054655TP, R1054655B, 1054655, 1054655TPV, 1054655TP, 1054655B

Why the product was recalled

The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.

Distribution information

US Nationwide and Global Distribution to countries of: Australia, Brazil, Canada, China, Germany, Japan, Korea, and Latin America.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.