Recall record at a glance
| Official record | Z-2625-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Remote Diagnostic Technologies Ltd. |
| Report date | Sep 27, 2023 |
| Recall initiation date | Aug 24, 2023 |
| Quantity | 3,524 |
| Location | Farnborough, United Kingdom |
Official product description
Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video Laryngoscope, REF 01-2044
Why the product was recalled
Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required. Error prevents: viewing graphical patient vital signs, sounding of pulse audio tone, visualization of patient airway used during intubation; may lead to treatment delay or hypoxia.
Distribution information
US:AZ, AL, CA, CO, DC, DE, FL , GA, HI, IA, ID, IL, IN , KS, KY, MA , MD , ME, MI, MN, MO, MS, MT, NC , ND, NE, NJ, NM , NV, NY, OH, OK, OR, PA, TN, TX, UT, VA , WA, WI, WV, WY. OUS: AE, AT, AU, BE, BN, BO, CA, CH, CZ, DE, DK, ES, FR, GB, HK, IE, IL, IS, IT, KY, LU, MY, NL, NO, OM, PT, RO, SA, SE, SG, SI, TH, YT, ZA
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.