FDA recall record

Philips SureSigns VM 3/4/6/8 Patient Monitors, 863063 863064 863065 863066 863068 863077 recall — FDA Z-2895-2018

Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or…

Class IIclassification
Z-2895-2018official record ID
Aug 29, 2018report date

Recall record at a glance

Official recordZ-2895-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPhilips Electronics North America Corporation
Report dateAug 29, 2018
Recall initiation dateJul 13, 2018
QuantityVM 96,324 Units; VS 110,424 Units; VSV 691 Units
LocationAndover, MA, United States

Official product description

Philips SureSigns VM 3/4/6/8 Patient Monitors, 863063 863064 863065 863066 863068 863077 863085 863086 863317 863287 863288 863289 The SureSigns VM4, VM6, and VM8 patient monitors are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Standard and optional parameters include: " ECG " Respiration " NBP " S2pO " IBP " CO2 " Temperature

Why the product was recalled

Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.

Distribution information

Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic, Dominican Republic, El Salvador, Egypt, Estonia, Ethiopia, France, French Guiana, French Polynesia, Finland, Guadeloupe, Germany, Ghana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Kenya, Korea, Kosovo, Republic of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Lybia, Macao, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, Zambia and Zimbabwe.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.