Philips HeartStart MRx, Monitor/Defibrillators with model numbers M3535A, M3536A, M3536M recall — FDA Z-2794-2020

The therapy switch may fail, resulting in the device exhibiting the following behaviors: the device may not perform the selected function, the therapy…

Class IIclassification
Z-2794-2020official record ID
Aug 19, 2020report date

Recall record at a glance

Official recordZ-2794-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPhilips North America, LLC
Report dateAug 19, 2020
Recall initiation dateJul 16, 2020
Quantity95,212 units
LocationAndover, MA, United States

Official product description

Philips HeartStart MRx, Monitor/Defibrillators with model numbers M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9 - Product Usage: is intended for use in hospital and pre-hospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support or defibrillation.

Why the product was recalled

The therapy switch may fail, resulting in the device exhibiting the following behaviors: the device may not perform the selected function, the therapy knob may not change to the energy setting selected, or the device may deliver a shock with an energy level different from the setting selected by the user. These device behaviors could result in a delay in therapy or a failure to deliver the intended therapy.

Distribution information

Worldwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceSep 9, 2026

Philips Brilliance recall — FDA Z-3040-2026

Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.

Class IIFDA: OngoingPhilips North America LLC