Philips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A recall — FDA Z-1923-2019

The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power…

Class IIclassification
Z-1923-2019official record ID
Jul 17, 2019report date

Recall record at a glance

Official recordZ-1923-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPhilips North America, LLC
Report dateJul 17, 2019
Recall initiation dateJun 7, 2019
Quantity97,794 Units
LocationAndover, MA, United States

Official product description

Philips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9

Why the product was recalled

The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery of a shock or pacing can occur. A faulty AC Power Module may also fail to charge the battery.

Distribution information

Nationwide domestic distribution. Worldwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceSep 9, 2026

Philips Brilliance recall — FDA Z-3040-2026

Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.

Class IIFDA: OngoingPhilips North America LLC