FDA recall record

Invivo Philips Expression MR200 MRI Patient Monitoring System Model 866120 recall — FDA Z-0001-2015

During production, excess metal (aluminum) debris was found stuck to the cling plastic that covers the device's pole assembly. The metallic debris could…

Class IIclassification
Z-0001-2015official record ID
Oct 8, 2014report date

Recall record at a glance

Official recordZ-0001-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmInvivo Corporation
Report dateOct 8, 2014
Recall initiation dateSep 8, 2014
Quantity10 units
LocationOrlando, FL, United States

Official product description

Philips Expression MR200 MRI Patient Monitoring System Model 866120

Why the product was recalled

During production, excess metal (aluminum) debris was found stuck to the cling plastic that covers the device's pole assembly. The metallic debris could affect system performance due to the potential for a failure to power up or loss of power. This could result in the loss of monitoring and a delay in diagnosis of the patient.

Distribution information

Worldwide Distribution: US distribution in state of: CA and countries of: Poland, Singapore, and Sweden.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.