Recall record at a glance
| Official record | Z-0944-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Philips Ultrasound Inc. |
| Report date | Feb 10, 2021 |
| Recall initiation date | Dec 16, 2020 |
| Quantity | 607 units |
| Location | Bothell, WA, United States |
Official product description
Philips EPIQ Diagnostic Ultrasound System; Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C, EPIQ CVx & EPIQ CVxi. MOD EPIQ Elite DIAGNOSTIC ULTRASOUND SYSTEM; Rx Only 100-240V~50/60 Hz, 600 VA MOD EPIQ 5 DIAGNOSTIC ULTRASOUND SYSTEM; GMDN 40761, Rx Only, 100-240V~50/60 Hz, 600VA MOD EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM; GMDN 40761, Rx Only, 100-240V~50/60 Hz, 600VA MOD EPIQ DIAGNOSTIC ULTRASOUND SYSTEM; Rx Only,100-240V~50/60 Hz, 600VA
Why the product was recalled
The manufacturer discovered an issue associated with the EPIQ Image Boost with xPlane and Color Flow or Doppler while using the X8-2t TEE Transducer. If Image Boost is enabled the color box shows flow visualization not accurately represent the fluid flow. Similarly, if Image Boost is enabled, both CW and PW Doppler traces will not accurately represent the fluid flow.
Distribution information
US - AL, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE,NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI and WV OUS- Austria, Australia, Belgium, Canada, Chezh Republic, Croatia, Denmark, Egypt, Finland, France, Hong Kong, Indonesia, India, Ireland, Italy, Japan, Latvia, Lithuania, Malta, Mexico, Netherlands, Norway, Poland, Romania, Russia, Senegal, Slovenia, Sri Lanka, South Africa, Switzerland, Thailand, Turkey, UAE and United Kingdom.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.