FDA recall record

Childrens Medical Ventures Philips/Children's Medical Ventures Smart Monitor 2 Professional recall — FDA Z-1789-2014

The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results…

Class IIclassification
Z-1789-2014official record ID
Jun 25, 2014report date

Recall record at a glance

Official recordZ-1789-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmChildrens Medical Ventures
Report dateJun 25, 2014
Recall initiation dateApr 29, 2014
Quantity112
LocationMonroeville, PA, United States

Official product description

Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.

Why the product was recalled

The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.

Distribution information

Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.