FDA recall record

Philips Azurion systems with software version R1.2 -Interventional Fluoroscopic X-ray system recall — FDA Z-2205-2019

The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result…

Class IIclassification
Z-2205-2019official record ID
Aug 14, 2019report date

Recall record at a glance

Official recordZ-2205-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPhilips North America, LLC
Report dateAug 14, 2019
Recall initiation dateJul 16, 2019
Quantity834
LocationAndover, MA, United States

Official product description

Philips Azurion systems with software version R1.2 -Interventional Fluoroscopic X-ray system as follows: Azurion 3 M12, Azurion 3 M15, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20 Models: 722063, 722064, 722067, 722068, 722078, 722079. Product Usage: The Azurion series (within the limits of the used Operating Room table) are intended for use to perform: Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: . The Azurion series can be used in a hybrid Operating Room. The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.

Why the product was recalled

The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.

Distribution information

US Nationwide Distribution Foreign: Canada Argentina AUSTRALIA Austria Azerbaijan Bahrain Bangladesh Belgium Bosnia and Herzegovina Brazil Bulgaria Chile Colombia CROATIA Denmark ECUADOR Egypt Finland France Germany Greece Hong Kong India Indonesia Ireland Israel Italie Italy Japan Jordan Kazakhstan Korea, Republic of Latvia Lebanon Libya Malaysia Maldives Mexico Mongolia Myanmar Netherlands New Zealand Norway Oman Pakistan Philippines Polamnd Poland Portugal Republic of Korea Romania Russian Federation Saudi Arabia Serbia Singapore SLOVENIA South Africa Spain Sri Lanka Sweden Switzerland Thailand Tunisia Turkey UNITED ARAB EMI United Arab Emirates United Kingdom Uzbekistan Viet Nam

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.