FDA recall record

Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1 recall — FDA Z-2284-2026

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user…

Class IIclassification
Z-2284-2026official record ID
Jun 10, 2026report date

Recall record at a glance

Official recordZ-2284-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Report dateJun 10, 2026
Recall initiation dateDec 7, 2023
Quantity16 units
LocationBest, Netherlands

Official product description

Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

Why the product was recalled

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.