Recall record at a glance
| Official record | Z-1144-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. |
| Report date | Feb 28, 2024 |
| Recall initiation date | Jan 2, 2024 |
| Quantity | 1,993 US; 2,662 OUS |
| Location | Best, N/A, Netherlands |
Official product description
Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of affected systems. Affected PCs: FlexVision PC Allura Xper Model Name - Model Number: Allura Xper FD10 722003 722010 722026; Allura Xper FD10/10 722005 722011 722027; Allura Xper FD20 722006 722012 722028; Allura Xper FD20 Biplane 722008 722013; Allura Xper FD20 Biplane OR Table 722025; Allura Xper FD20 OR Table 722023 722035; Allura Xper FD20/10 722029; Allura Xper FD20/15 722058; Allura Xper FD20/15 OR Table 722059; Allura Xper FD20/20 722038; Allura Xper FD20/20 OR Table 722039;
Why the product was recalled
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
Distribution information
Worldwide - US Nationwide distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.