Philips Allura System configured with a 1 Phase UPS recall — FDA Z-2503-2025

Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting…

Class IIclassification
Z-2503-2025official record ID
Sep 10, 2025report date

Recall record at a glance

Official recordZ-2503-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Report dateSep 10, 2025
Recall initiation dateAug 1, 2025
Quantity1,006 units
LocationBest, N/A, Netherlands

Official product description

Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.

Why the product was recalled

Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.

Distribution information

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Latvia, Lithuania, Macedonia, Maldives, Malta, Mexico, Moldova, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Panama, Paraguay, Peru, Philippines, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.