FDA recall record

PHADIA Varelisa Cardiolipin IgM Cardiolipin autoantibody immunological test system recall — FDA Z-2285-2012

Customers were not following the labeled assay procedure. This was due to a letter issued to them in 2004 from Phadia GmbH, which indicated that customers…

Class IIIclassification
Z-2285-2012official record ID
Sep 5, 2012report date

Recall record at a glance

Official recordZ-2285-2012
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmPhadia US Inc.
Report dateSep 5, 2012
Recall initiation dateAug 2, 2012
Quantity837 kits (96 assays per kit), total of 80,352 individual assays
LocationPortage, MI, United States

Official product description

PHADIA Varelisa Cardiolipin IgM Cardiolipin autoantibody immunological test system (IgM class antibodies)

Why the product was recalled

Customers were not following the labeled assay procedure. This was due to a letter issued to them in 2004 from Phadia GmbH, which indicated that customers could omit the pre-wash step. The omission of the pre-wash step, when combined with a change in material source in kit lot #0067, caused elevated results.

Distribution information

USA Nationwide.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.