Recall record at a glance
| Official record | Z-1224-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Zimmer Biomet, Inc. |
| Report date | Apr 4, 2018 |
| Recall initiation date | Dec 12, 2017 |
| Quantity | 3 units |
| Location | Warsaw, IN, United States |
Official product description
PERSONA THE PERSONALIZED KNEE SYSTEM "PSN TIB STM 5 DEG SZ D R" indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
Why the product was recalled
A complaint was received for Item 42-5320-067-02, lot 63236047. It was report that there was no set screw or taper plug included in the box. Complaint investigation confirmed that the entire lot of 63236047 was missing the set screw and taper plug.
Distribution information
Non-USA Customer
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.