Recall record at a glance
| Official record | Z-1758-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | SPINEART SA |
| Report date | May 15, 2024 |
| Recall initiation date | Mar 15, 2024 |
| Quantity | 242 systems |
| Location | Plan-Les-Ouates, N/A, Switzerland |
Official product description
PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (painful degeneration of the disc), spondylolisthesis, trauma, spinal stenosis, deformities (i.e. scoliosis, kyphosis, or lordosis), tumor and failed previous fusion (pseudarthrosis).
Why the product was recalled
Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew
Distribution information
U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.