| Official record | Z-2530-2016 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Danville Materials, Inc. |
| Report date | Aug 24, 2016 |
| Recall initiation date | Jun 24, 2016 |
| Quantity | ~ 2123 units |
| Location | San Ramon, CA, United States |
Official product description
Perioscopy Sheath: Part number: 1008; The Perioscopy Sheath is a single-use sterile component used with the Perioscopy System, which is a medical device that features miniaturized digital video technology, and enables the healthcare professional to see magnified details of the tooth anatomy and deposits below the gum line (i.e. calculus root fractures, restoration margins, tooth decay).
Why the product was recalled
The Perioscopy Sheaths are currently provided with a label that indicates the Sheath has a 3 year shelf-life/expiry date. Danville Materials has become aware that validation data are not available to support the 3-year shelf-life/expiration dating claim.
Distribution information
US and New Zealand, Australia, Canada, Japan, Israel, United Kingdom, Italy, Mexico.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.