Teleflex Medical Percuvance Johans Grasper Tool Tip The Percutaneous Surgical System with recall — FDA Z-3221-2017

For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots…

Class IIclassification
Z-3221-2017official record ID
Oct 4, 2017report date

Recall record at a glance

Official recordZ-3221-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTeleflex Medical
Report dateOct 4, 2017
Recall initiation dateJul 19, 2017
Quantity20,039 (US and OUS)
LocationMorrisville, NC, United States

Official product description

Percuvance Johans Grasper Tool Tip The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.

Why the product was recalled

For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.

Distribution information

Worldwide Distribution - US including AL, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV Internationally to United Arab Emirates, Vienna, Australia, Belgium, Switzerland, China, Germany, Czech Republic, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, Iran, Italy, Japan, South Korea, R¿publique du Liban, Netherlands, Norway, Qatar, Romania, Saudi Arabia, Republic of Singapore, Turkey, Taiwan, and South Africa

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.