FDA recall record

Oculus Optikgeraete GMBH Pentacam AXL REF 70100, Pentacam HR REF 70900 recall — FDA Z-1673-2022

Due to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs

Class IIclassification
Z-1673-2022official record ID
Sep 7, 2022report date

Recall record at a glance

Official recordZ-1673-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOculus Optikgeraete GMBH
Report dateSep 7, 2022
Recall initiation dateJul 8, 2022
Quantity92 systems with affected software
LocationWetzlar, Germany

Official product description

Pentacam AXL REF 70100, Pentacam HR REF 70900. Used to image the anterior segment of the eye.

Why the product was recalled

Due to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs

Distribution information

U.S.: AL, AR, AZ, CAQ, CO, CT, FL, GA, HI, IA, ID, IL, KS, KY, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WY O.U.S.: N/A

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records