FDA recall record

Zien Medical Technologies PenBlade Safety Scalpel, for the incision of human skin and recall — FDA Z-2800-2015

Cracks were discovered in the PET tray of a small population of sterile product. These cracks create a risk of loss of sterile barrier for the product.…

Class IIclassification
Z-2800-2015official record ID
Sep 30, 2015report date

Recall record at a glance

Official recordZ-2800-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmZien Medical Technologies, Inc.
Report dateSep 30, 2015
Recall initiation dateAug 18, 2015
Quantity28,284
LocationSalt Lake City, UT, United States

Official product description

PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-M-11P, PB-M-15

Why the product was recalled

Cracks were discovered in the PET tray of a small population of sterile product. These cracks create a risk of loss of sterile barrier for the product. Sampling and root cause analysis indicated that the risk could extend throughout the lots manufactured from September 2014-July 2015. This led to the decision to recall the product.

Distribution information

Worldwide distribution to U.S. nationwide, Germany and England.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.