FDA recall record

Edwards Lifesciences PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM recall — FDA Z-2479-2023

Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced…

Class IIclassification
Z-2479-2023official record ID
Sep 6, 2023report date

Recall record at a glance

Official recordZ-2479-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmEdwards Lifesciences, LLC
Report dateSep 6, 2023
Recall initiation dateJul 27, 2023
Quantity92
LocationIrvine, CA, United States

Official product description

PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM and 20000IS

Why the product was recalled

Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system by rotating the release knob. Additional torque required to initiate rotation of the release wire from the implant, could cause implant rotation or additional implant implantation.

Distribution information

International distribution to the countries of Austria, Belgium, Switzerland, Germany, Cyprus, Israel, UK, Greece, Ireland, Italy, Luxembourg, Poland.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.