FDA recall record

Philips PageWriter TC30 w/o trolley Government Bundle, Product 860355, Software revisions up recall — FDA Z-0861-2019

Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can…

Class IIclassification
Z-0861-2019official record ID
Feb 27, 2019report date

Recall record at a glance

Official recordZ-0861-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPhilips North America, LLC
Report dateFeb 27, 2019
Recall initiation dateJan 9, 2019
Quantity14
LocationAndover, MA, United States

Official product description

PageWriter TC30 w/o trolley Government Bundle, Product 860355, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients

Why the product was recalled

Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.

Distribution information

Domestic distribution nationwide. Foreign distribution worldwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceSep 9, 2026

Philips Brilliance recall — FDA Z-3040-2026

Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.

Class IIFDA: OngoingPhilips North America LLC