FDA recall record

Fisher Diagnostics Pacific Hemostasis FDP Latex Reagent, REF/Model Number 100651, UDI recall — FDA Z-2763-2017

An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list…

Class IIclassification
Z-2763-2017official record ID
Aug 2, 2017report date

Recall record at a glance

Official recordZ-2763-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmFisher Diagnostics
Report dateAug 2, 2017
Recall initiation dateMay 19, 2017
Quantity27 units
LocationMiddletown, VA, United States

Official product description

Pacific Hemostasis FDP Latex Reagent, REF/Model Number 100651, UDI 00845275000870, 1 x 5 mL, IVD; --- Fisher Diagnostics, a division of Fisher Scientific Company, LLC, a part of Thermo Fisher Scientific Inc., Middletown, VA 22645-1905 USA

Why the product was recalled

An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs accurately at FDP concentrations greater than 20 ¿g/ml, there is an observed reduction in sensitivity when FDP levels are less than 20 ¿g/ml. This reduction in sensitivity has been tied to FDP Latex Reagent lot 890199 and may result in false negatives.

Distribution information

USA (nationwide) including Puerto Rico, and Internationally to Colombia

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.