FDA recall record

Trividia Health Owner's Booklets and Instructions for Use that are used with the following recall — FDA Z-1545-2026

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code…

Class Iclassification
Z-1545-2026official record ID
Mar 18, 2026report date

Recall record at a glance

Official recordZ-1545-2026
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmTrividia Health, Inc.
Report dateMar 18, 2026
Recall initiation dateFeb 6, 2026
Quantity12,067,458
LocationFort Lauderdale, FL, United States

Official product description

Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX Kit/RE4001-01, Leader TRUE METRIX Kit/RE4002-01, Leader TRUE METRIX Meter Only/RE4002-40, CVS TRUE METRIX Kit/RE4007-01SB, GNP TRUE METRIX Meter Only/RE4011-00, GNP TRUE METRIX Kit /RE4011-01, MEIJER TRUE METRIX Kit/RE4019-01, FRED'S TRUE METRIX Kit/RE4023-01, DISCOUNT DRUG TRUE METRIX Kit/RE4025-01, KINRAY TRUE METRIX Kit/RE4027-01, HEB TRUE METRIX Kit/RE4031-01, McKesson TRUE METRIX Starter Kit/RE4051-43, SUNMARK TRUE METRIX Meter Only/RE4054-00, SUNMARK TRUE METRIX Kit/RE4054-01, Signature Care TRUE METRIX Kit/RE4060-01, CareOne TRUE METRIX Kit/RE4061-01, RITE AID TRUE METRIX Kit/RE4066-01, Liberty TRUE METRIX Meter Only/RE4078-40, HUMANA TRUE METRIX Starter Kit/RE4081-03, Wegmans TRUE METRIX Kit/RE4087-01, HEALTHMART TRUE METRIX Meter Only/RE4089-00, HEALTHMART TRUE METRIX Kit/RE4089-01, ARRIVA TRUE METRIX Meter Only/RE4095-40, HealthyAccents TRUE METRIX Kit/RE4097-01, Shopko TRUE METRIX Kit/RE4098-01, Henry Schein TRUE METRIX Kit/RE4099-40, Henry Schein TRUE METRIX Meter Only/RE4099-45, Millennium Pharmay Services TRUE METRIX Kit/RE4202-43, Kroger TRUE METRIX Kit/RE4203-01, TOPCARE TRUE METRIX Kit/RE4209-01, Foster and Thrive TRUE METRIX Kit/RE4211-01, TRUE METRIX Meter Only/RE4H01-00, TRUE METRIX Kit/RE4H01-01, TRUE METRIX Starter Kit/RE4H01-04, TRUE METRIX NFRS Meter Only/RE4H01-40, TRUE METRIX NFRS Starter Kit/RE4H01-43, TRUE METRIX Kit/RE4H01-51, TRUE METRIX Meter Only (mg/dL)/RE4i01-00, TRUE METRIX Kit (mg/dL)/RE4i01-01, TRUE METRIX Kit (mg/dL)/RE4i01-02, TRUE METRIX Starter Kit (mg/dL)/RE4i03-02, TRUE METRIX Kit (mmol/L)/RE4i03-11, Farmacia Benavides TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i05-00, Farmacia del Ahorro TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i07-00, TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i23-00, TRUE METRIX Meter Only (Jamaica - mmol/L)/RE4i29-11, TRUE METRIX Kit (mg/dL)/RE4i61-02, TRUE METRIX Kit (mg/dL)/RE4i61-02THI, TRUE METRIX Kit (United Kingdom - mmol/L)/RE4i82-11, TRUE METRIX Starter Kit (United Kingdom - mmol/L)/RE4i82-12.

Why the product was recalled

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

Distribution information

Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.