FDA recall record

Palodex Group Oy Orthopantomograph OP300 dental computed tomography x-ray system recall — FDA Z-2413-2024

It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result…

Class IIclassification
Z-2413-2024official record ID
Aug 21, 2024report date

Recall record at a glance

Official recordZ-2413-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPalodex Group Oy
Report dateAug 21, 2024
Recall initiation dateOct 28, 2021
Quantity103
LocationTuusula, Finland

Official product description

Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.

Why the product was recalled

It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).

Distribution information

US Nationwide-Worldwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: FDA Mandated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.