FDA recall record

Orthofix MINIRAIL, SHORT, LENGTHENER REF: M103, Unilateral orthopedic/trauma external recall — FDA Z-2730-2026

Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122…

Class IIclassification
Z-2730-2026official record ID
Jul 22, 2026report date

Recall record at a glance

Official recordZ-2730-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOrthofix U.S. LLC
Report dateJul 22, 2026
Recall initiation dateJun 10, 2026
Quantity19 units
LocationLewisville, TX, United States

Official product description

Orthofix MINIRAIL, SHORT, LENGTHENER REF: M103, Unilateral orthopedic/trauma external fixation device

Why the product was recalled

Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111

Distribution information

US: AZ CA CO CT FL GA ID IL IN LA MD MI MO MS MT NC NY OH OR SC SD TX UT VA WA WI OUS: None

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.