FDA recall record

LEONI CIA CABLE SYSTEMS ORION System with software OSS v6.0 and OSS v7.0- An recall — FDA Z-0399-2024

Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when…

Class IIclassification
Z-0399-2024official record ID
Nov 29, 2023report date

Recall record at a glance

Official recordZ-0399-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmLEONI CIA CABLE SYSTEMS
Report dateNov 29, 2023
Recall initiation dateOct 6, 2023
Quantity3 units
LocationChartres, N/A, France

Official product description

ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00009

Why the product was recalled

Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.

Distribution information

US Nationwide distribution in the states of MA, TN.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.