FDA recall record

OriGen BIOMEDICAL VV13F Reinforced Dual Lumen Catheter STERILE DISPOSABLE: DO NOT RE-USE recall — FDA Z-1456-2015

Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent…

Class Iclassification
Z-1456-2015official record ID
Apr 29, 2015report date

Recall record at a glance

Official recordZ-1456-2015
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmOriGen Biomedical, Inc.
Report dateApr 29, 2015
Recall initiation dateMar 30, 2015
Quantity51 units
LocationAustin, TX, United States

Official product description

OriGen BIOMEDICAL VV13F Reinforced Dual Lumen Catheter STERILE DISPOSABLE: DO NOT RE-USE Manufactured by: OriGen Biomedical. OriGene VV13F Reinforced Dual Lumen Catheter indicated for use as a single cannula for both venous drainage and re-infusion of blood in the internal jugular vein during extracorporeal life support procedures of six hours or less.

Why the product was recalled

Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage.

Distribution information

Worldwide Distribution: US (nationwide) and countries of: Spain, UK, Germany, Sweden, and Netherlands.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.