FDA recall record

OraSure Technologies OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807 recall — FDA Z-2403-2016

Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C…

Class Iclassification
Z-2403-2016official record ID
Aug 24, 2016report date

Recall record at a glance

Official recordZ-2403-2016
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmOraSure Technologies, Inc.
Report dateAug 24, 2016
Recall initiation dateJul 20, 2016
Quantity3,075 units
LocationBethlehem, PA, United States

Official product description

OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.

Why the product was recalled

Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.

Distribution information

Distributed to one customer only in US state of Georgia.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records