Alcon Research ORA System with VerifEye, Catalog Number 8065998300 For use during recall — FDA Z-3050-2017

Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a…

Class IIclassification
Z-3050-2017official record ID
Aug 30, 2017report date

Recall record at a glance

Official recordZ-3050-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAlcon Research, Ltd.
Report dateAug 30, 2017
Recall initiation dateJun 30, 2017
Quantity367 units
LocationFort Worth, TX, United States

Official product description

ORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgery

Why the product was recalled

Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.

Distribution information

Worldwide distribution. US Nationwide and Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, France, India, Japan, Mexico, Netherlands, Panama, Portugal, Romania, Spain, Thailand, UAE, and United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 23, 2023

Alcon Custom Pak recall — FDA Z-2411-2023

Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

Class IIFDA: OngoingAlcon Research, LTD.