FDA recall record

Argon Medical Devices Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code recall — FDA Z-0589-2026

Due to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays…

Class IIclassification
Z-0589-2026official record ID
Dec 3, 2025report date

Recall record at a glance

Official recordZ-0589-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmArgon Medical Devices, Inc.
Report dateDec 3, 2025
Recall initiation dateSep 12, 2025
QuantityTotal devices 1821 (US=1794 and O.U.S.=27) units
LocationAthens, TX, United States

Official product description

Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated

Why the product was recalled

Due to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays and/or risk of venous intimal injuries if excessive force is applied

Distribution information

U.S. (nationwide) distribution to states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. O.U.S. (international) to countries of: China, Panama and Trinidad and Tobago

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.