The Binding Site Group Optilite IgG4 Kit Product Code: LK009.OPT.A recall — FDA Z-1089-2017

A change to the lower end of the measuring range and product insert did not reflect the change.The measuring range provided in instrument parameters is…

Class IIIclassification
Z-1089-2017official record ID
Feb 1, 2017report date

Recall record at a glance

Official recordZ-1089-2017
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmThe Binding Site Group, Ltd.
Report dateFeb 1, 2017
Recall initiation dateDec 23, 2016
Quantity9 users
LocationBirmingham, N/A, United Kingdom

Official product description

Optilite IgG4 Kit Product Code: LK009.OPT.A

Why the product was recalled

A change to the lower end of the measuring range and product insert did not reflect the change.The measuring range provided in instrument parameters is correct but contradicts the information provided in the product insert

Distribution information

U.S. distribution to the following; FL, TX, PA, CA, MA, WA, MI, CT. No foreign distribution.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.