FDA recall record

OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004 recall — FDA Z-2310-2021

The firm identified a precipitation issue which may cause late Ct reactors (a false positive result) with a rate of 5 to 10% for affected vials of RNA…

Class IIclassification
Z-2310-2021official record ID
Aug 25, 2021report date

Recall record at a glance

Official recordZ-2310-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOPTI Medical Systems, Inc.
Report dateAug 25, 2021
Recall initiation dateJun 25, 2021
Quantity3552 kits
LocationRoswell, GA, United States

Official product description

OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004

Why the product was recalled

The firm identified a precipitation issue which may cause late Ct reactors (a false positive result) with a rate of 5 to 10% for affected vials of RNA Master Mix.

Distribution information

Maine, New York, Alabama, Pennsylvania, Florida, Illinois, New Jersey, Connecticut, Georgia, Puerto Rico, Costa Rica, Hong Kong, Philippines, Barbados

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.