Nidek OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 recall — FDA Z-2477-2015

Software bug was found where there was no difference in Total and Corneal high-order aberrations, but differences were found in Internal high-order…

Class IIclassification
Z-2477-2015official record ID
Sep 2, 2015report date

Recall record at a glance

Official recordZ-2477-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNidek Inc.
Report dateSep 2, 2015
Recall initiation dateJun 15, 2015
Quantity3836
LocationFremont, CA, United States

Official product description

OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01. Opthalmic: The OPD-Scan III is a diagnostic instrument that is indicated for use for Mapping of refractive error distribution of the eye by measurement and analysis of spherical power, cylindrical power, and cylinder axis.

Why the product was recalled

Software bug was found where there was no difference in Total and Corneal high-order aberrations, but differences were found in Internal high-order aberrations.

Distribution information

Nationwide Distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.